Post market vigilance (pmv) received one device.The visual inspection of the returned product noted that the circular seal was cut.The trocar, cannula and envelope seal appeared intact.The obturator was received.The condition of the device precludes functional testing.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Analysis concluded there were no assembly or component related failures.Replication of the cut in the circular seal may occur when mishandled during clinical application.The root cause of the observed damage was misuse of the product which would have caused or contributed to the reported incident.Should new information become available, the file will be re-opened and the investigation summary will be amended as appropriate.If information is provided in the future, a supplemental report will be issued.
|