| Model Number |
MNC25015 |
| Medical Device Problem Codes |
Sticking (1597); Device Issue (2379); Physical Resistance (2578); Physical Property Issue (3008)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
04/10/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Review of device history records including raw material inspection, in process and finished product inspection does not reveal any discrepancy relevant to batch under investigation which confirms that there is no indication of a product related quality deficiency associated to this batch.Control sample test results of the same manufacturing lot again confirmed that there was no indication of a product quality deficiency.Additionally, a review of the complaint history identified no other incidents from this lot.
|
| |
|
Event or Problem Description
|
|
The mozec 2.5 x 15 mm nc was used to post dilate while on asahi 0.014" fielder xt-a wire.The balloon was inflated to 20 rbp and it appeared to be working well.After deflating mozec balloon, fellow dr.(b)(6) noticed it was slightly difficult to move back.Was able to get balloon and wire back into guide catheter.Then, had to clamp the wire to get mozec nc balloon off of the asahi fielder xt-a wire.
|
| |
|