The patient was undergoing a thrombectomy procedure using a velocity delivery microcatheter (velocity).During the procedure, the physician was unable to advance a guidewire through the velocity after advancing the velocity into the patient using a non-penumbra catheter.The velocity was therefore removed.Upon removal, the physician noticed that the velocity was kinked.The procedure was completed using a new velocity and the same catheter.There was no report of an adverse effect to the patient.
|
Results: the velocity was fractured approximately 67.0 cm from the hub.The velocity also had a slight bend on the strain relief near the hub.A non-penumbra guidewire was found fractured within the velocity.The velocity was damaged and, therefore, was not able to be functionally tested.Conclusions: evaluation of the returned velocity revealed a fractured device.The velocity was reportedly advanced into the non-penumbra catheter without a guidewire.If the catheter is already placed in anatomy and a guide wire is attempted to be advanced through the catheter, resistance may be experienced.If the guidewire is forcefully manipulated against this resistance, the guidewire will likely damage the catheter and damage such as a kink may occur.Further evaluation revealed that the device was fractured and bent.This damage was not mentioned in the reported complaint, therefore the fracture and bend are likely incidental and may have occurred during removal from the patient or during packaging for return to penumbra.Penumbra catheters are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|