Catalog Number RBYPOD5
Medical Device Problem Codes Failure to Advance (2524); Physical Resistance (2578)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 04/10/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Results: the embolization coil was returned detached from its pusher assembly with the proximal constraint sphere intact.Conclusions: evaluation of the returned pod5 revealed that the embolization coil was detached from its pusher assembly.However, the pod5 pusher assembly, introducer sheath and the px slim referred to in the complaint description were not returned to penumbra for evaluation.Therefore, the root cause of the reported resistance could not be determined.Penumbra catheters and coils are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure using pod5s.During the procedure, the physician felt resistance and was unable to advance a pod5 out of its introducer sheath and into a px slim delivery microcatheter (px slim).The pod5 was therefore removed, and the procedure was completed using the same px slim and a pod packing coil (podj).There was no report of an adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POD5
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key7491783
Report Number3005168196-2018-00961
Device Sequence Number18915597
Product Code HCG
UDI-Device Identifier00814548013787
UDI-Public00814548013787
Combination Product (Y/N)N
Initial Reporter CountryHK
PMA/510(K) Number
K141134
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign,health pr
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 04/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date06/11/2025
Device Catalogue NumberRBYPOD5
Device Lot NumberF77643
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/24/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/11/2018
Initial Report FDA Received Date05/07/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age68 YR
-
-