| Model Number |
TRUETRACK |
| Medical Device Problem Code |
Device Displays Incorrect Message (2591)
|
| Health Effect - Clinical Code |
Shaking/Tremors (2515)
|
| Date of Event |
04/13/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Product not returned for evaluation.No replacement product needed at this time.Complaint resolved by training during the time of the call.Product is working as designed.Most likely underlying root cause: mlc-68-user did not insert code key test strip udi# (b)(4).Note: manufacturer contacted customer (several attempts) in a follow-up call in order to ensure the customer's condition since the initial call - unable to establish contact with the customer at this time.Product notification letter sent to customer to contact customer care.
|
| |
|
Event or Problem Description
|
|
Consumer reported complaint for code (----) eci meter.The meter read code (----) as soon as the test strip was inserted.The code chip matched the test strip.Customer did not have a new vial of test strips.During the call on (b)(6) 2018, a blood test was performed by the customer fasting after reinserting the code chip and produced test result of 89 mg/dl using truetrack meter.The customer's expected fasting blood glucose test result range was undisclosed.At the time of the call the customer reported feeling shaky; medical attention was not needed at the time of the call.The product storage was undisclosed.The test strip lot manufacturer's expiration date is 03/31/2019 and open vial date was undisclosed.The meter memory was not reviewed for previous test result history.Complaint resolved during the call.
|
| |
|