• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: JOHNSON & JOHNSON SURGICAL VISION, INC. TECNIS; TORIC IOLS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

JOHNSON & JOHNSON SURGICAL VISION, INC. TECNIS; TORIC IOLS Back to Search Results
Model Number ZCT150
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Capsular Bag Tear (2639); Vitrectomy (2643)
Date of Event 04/03/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If implanted; give date: n/a (not applicable).The intraocular lens was removed and replaced during the same procedure.If explanted; give date: n/a (not applicable).The intraocular lens was removed and replaced during the same procedure.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that after the zct150 22.0 diopter intraocular lens (iol) was inserted, a capsule tear occurred.The doctor performed vitrectomy.The lens was removed and reportedly it was replaced with a sulcus 3 piece lens.There were no other interventions required and the patient outcome at the time of discharge was good.There was no patient injury reported.No additional information was provided.
 
Additional Manufacturer Narrative
Device available for evaluation? yes.Returned to manufacturer on: 5/17/2018.Device returned to manufacturer? yes.Device evaluation: the device was received at the manufacturing site for investigation.Visual inspection shows there was cut on the lens most probably to make remove and replacement possible.The lens is dirty.This is expected as the lens was transported from controlled environment during manufacturing to a non-sterile, non-environmental controlled area.Investigation of the return sample and the condition of the return sample do not suggest the dirt and cut was introduced during manufacturing.The complaint cannot be confirmed.Manufacturing record review: manufacturing record review of the production order and related document revealed that the product was manufactured and released according to specification.Historical data analysis: a search in the complaint history was performed on (b)(6) 2018 revealed that no similar complaints were received for this production order.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions for the proper use and handling of the product.Conclusion: as a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS
Common Device Name
TORIC IOLS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
santa ana CA 92705
MDR Report Key7492234
Report Number9614546-2018-00437
Device Sequence Number18915613
Product Code HQL
UDI-Device Identifier05050474552180
UDI-Public(01)05050474552180(17)210508
Combination Product (Y/N)Y
Initial Reporter StateLA
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/04/2018
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Physician
Device Expiration Date05/08/2021
Device Model NumberZCT150
Device Catalogue NumberZCT150U220
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/17/2018
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 04/10/2018
Supplement Date Received by Manufacturer08/21/2018
Initial Report FDA Received Date05/07/2018
Supplement Report FDA Received Date09/04/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
1MTEC30 CARTRIDGE, LOT#UNKNOWN; 1MTEC30 CARTRIDGE, LOT#UNKNOWN
Outcome Attributed to Adverse Event Required Intervention;
Patient Age73 YR
Date Report Sent to FDA01/01/2005
-
-