Model Number ENVEOR-N-US
Medical Device Problem Codes Positioning Failure (1158); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 10/24/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device history record was reviewed and showed that this product met all manufacturing specifications for product released for distribution.No issues were identified that would have impacted this event.Capsule separation typically occurred due to excessive compressive forces applied during the valve loading process.This excessive compressive force was typically experienced when the valve was loaded incorrectly.An assignable root cause related to this event could not be determined.The subject device was not returned for evaluation, and the reported capsule separation could not be confirmed.The valve and dcs were recaptured and removed from the patient.A new valve was loaded onto a new dcs and the valve was successfully implanted.No adverse patient effects were reported.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received information that during the implant of this transcatheter bioprosthetic valve, the delivery catheter system (dcs) was advanced, crossed the valve, and upon deployment the capsule did not move.It was reported that the capsule was broken and would not open.The valve and dcs were recaptured and withdrawn from the body.A new valve was loaded onto a new dcs and the valve was successfully implanted.No adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENVEO R DELIVERY SYSTEM
Common Device Name
AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED
Manufacturer (Section D)
MEDTRONIC HEART VALVES DIVISION
1851 e deere ave
santa ana CA 92705
MDR Report Key7492564
Report Number2025587-2018-01075
Device Sequence Number145171
Product Code NPT
UDI-Device Identifier00643169640337
UDI-Public00643169640337
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/20/2018
Device Model NumberENVEOR-N-US
Device Catalogue NumberENVEOR-N-US
Device Lot Number0008459338
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/01/2018
Initial Report FDA Received Date05/07/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age88 YR
-
-