| Model Number |
ASCA5001-01 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
04/13/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(6).
|
| |
|
Event or Problem Description
|
|
It was reported that the wand was put under no more stress than normal and part of the electrode popped off.
|
| |
|
Additional Manufacturer Narrative
|
|
The returned device, intended for use in treatment, was returned for evaluation.There was a relationship found between the returned device and the reported incident.Visual inspection under magnification shows extensive electrode wear and scratch/scuff marks on the spacer and cap.The screen is completely detached and returned with the device.A part of the spacer is broken-off.The insulation around the shaft is scratched but not compromised.There are no manufacturing abnormalities visually observed with the returned device.The device was connected to a compatible quantum2 controller and did not work; registered an e-7 error message.The e-7 error was bypassed and was activated in saline solution at the default and max settings on the controller and performed as intended.The complaint was verified as an e-7 error was experienced.The root cause could not be determined with confidence.The suction line was tested and performed as intended.
|
| |
|