| Model Number |
E SERIES |
| Medical Device Problem Code |
Device Displays Incorrect Message (2591)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has not received the device for evaluation and this complaint is still under investigation.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during a routine shift check by a clinician, the device displayed "defib fault 72" and "pacer fault 116" messages.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to (b)(6) medical corporation; the reported malfunction was observed and attributed to a system interconnect flex cable that was loose.A system interconnect flex cable and interlock connector were replaced to remedy the problem.Analysis for reports of this type has not identified an increase in trend.
|
| |
|