• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC IRELAND VALIANT CAPTIVIA; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC IRELAND VALIANT CAPTIVIA; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT Back to Search Results
Model Number VAMF3838C100TU
Medical Device Problem Code Difficult To Position (1467)
Health Effect - Clinical Code Perforation of Vessels (2135)
Date of Event 04/11/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
A valiant stent graft system was planned for implantation in a patient for the endovascular treatment of a 50 mm diameter a penetrating descending aortic aneurysm.An endurant stent graft system was also used for the treatment of an abdominal aortic aneurysm.The patients access vessels ranged in diameter between 6.2-7.2 mm.It was reported that during the index procedure the valiant device was attempted to be delivered but was unable to pass due to the narrow and calcified access vessels.The valiant was tried bare-back with no success.18 and 20f dilators were then used to attempt delivery of device again via an aortic bifurcate but this was unsuccessful.Angiography then showed perforation of the external iliac and distal common femoral arteries.This perforation required surgical bypass repair and additionally the aneurysm was successfully treated.The pau remains untreated.As per the physician, the cause of the event was anatomy related due to the narrow and calcified access vessels.No additional clinical sequelae were reported and the patient will be monitored.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VALIANT CAPTIVIA
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
MDR Report Key7494084
Report Number2953200-2018-00663
Device Sequence Number194158
Product Code MIH
UDI-Device Identifier00643169347045
UDI-Public00643169347045
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
P100040
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/05/2019
Device Model NumberVAMF3838C100TU
Device Catalogue NumberVAMF3838C100TU
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/11/2018
Initial Report FDA Received Date05/08/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age72 YR
-
-