| Model Number |
6DCFS |
| Medical Device Problem Code |
Human-Device Interface Problem (2949)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, the snap lock on the inner cannula broke off.Customer reported the patient was continuing to use the same trach and there was no re-cannulation or harm.And they have switched over to another inner cannula.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: sample was received and a visual inspection was conducted and it was observed that the flange is separated from the tube and was not received for evaluation; which would have been detected immediately in the manufacturing process since that defect is extremely obvious; therefore, the failure mode is not confirmed as related with the manufacturing process.Additionally a dimensional test was performed to the outer cannula holes and all the readings are within according to specification.If information is provided in the future, a supplemental report will be issued.
|
| |
|