• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER SUSTAINABILITY SOLUTIONS LAKELAND; SCALPEL, ULTRASONIC, REPROCESSED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER SUSTAINABILITY SOLUTIONS LAKELAND; SCALPEL, ULTRASONIC, REPROCESSED Back to Search Results
Model Number HARH36
Device Problem Device Inoperable (1663)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/21/2018
Event Type  malfunction  
Manufacturer Narrative
The device was returned to stryker sustainability solutions for evaluation.Visual inspection revealed evidence of clinical use.The blade was identified to be fractured and detached from the device.The detached blade was returned with the complaint device.The jaw, teflon pad and contact rings appear to be intact.A review of the dhr supports that the device met all inspection and test criteria prior to release from stryker.Therefore, the most likely root causes are: - jaws/blade subassembly damage - incidental and prolonged activation against solid surfaces, such as bone or plastic the instructions for use (ifu) state: - use the torque wrench (already mounted to the shaft) to tighten the blade onto the hand piece.Turn the torque wrench clockwise while holding only the gray hand piece until it clicks twice indicating that sufficient torque has been applied to secure the blade.- do not attempt to bend, sharpen, or otherwise alter the shape of the blade.Doing so may cause blade failure and user or patient injury.- avoid contact with any and all other instruments while the instrument is activated.Contact with staples, clips or other instruments while the instrument is activated may result in cracked or broken blades.- note: do not use any other means than the torque wrench to attach or detach the instrument from the hand piece.- note: do not torque the instrument by hand without the torque wrench or damage may occur to the hand piece.The reported event will continue to be monitored through post-market surveillance.
 
Event Description
It was reported the harmonic scalpel worked initially, then stopped working.There was no patient injury, medical intervention, or extended procedure time reported.These are commonly used devices that are readily available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NA
Type of Device
SCALPEL, ULTRASONIC, REPROCESSED
Manufacturer (Section D)
STRYKER SUSTAINABILITY SOLUTIONS LAKELAND
5307 great oak drive
lakeland FL 33815
Manufacturer (Section G)
STRYKER SUSTAINABILITY SOLUTIONS LAKELAND
5307 great oak drive
lakeland FL 33815
Manufacturer Contact
marissa richmond
1810 w. drake drive
tempe, AZ 85283
8888883433
MDR Report Key7494485
MDR Text Key107721471
Report Number0001056128-2018-00027
Device Sequence Number1
Product Code NLQ
UDI-Device Identifier07613327346251
UDI-Public(01)07613327346251(17)210221(10)8129034
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K161693
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 05/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/08/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/21/2021
Device Model NumberHARH36
Device Catalogue NumberHARH36RR
Device Lot Number8129034
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/02/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/11/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/20/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? Yes
Type of Device Usage Reuse
Patient Sequence Number1
-
-