| Model Number |
030453 |
| Medical Device Problem Code |
Difficult to Open or Close (2921)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
04/24/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, when the device was being applied to mesoappendix on a laparoscopic appendectomy, the staple load would not completely close on the tissue.They used a new staple cartridge and it worked fine.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.Single use loading unit opened by the account with the appropriate packaging in an undamaged condition.The anvil clamping mechanism was deformed.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Analysis concluded there were no assembly component related failures.The root cause of the observed damage was misuse of the product which would have caused or contributed to the reported incident.Replication of the anvil clamping feature deformation may occur through manipulation of tissue using the anvil side of the loading unit, or clamping over excessive thickness.If information is provided in the future, a supplemental report will be issued.
|
| |
|