• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE PKG, 5MM PEEK MONOPOLAR HANDLE; ENDOSCOPIC GRASPING/CUTTING INSTRUMENT, NON-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE PKG, 5MM PEEK MONOPOLAR HANDLE; ENDOSCOPIC GRASPING/CUTTING INSTRUMENT, NON-POWERED Back to Search Results
Catalog Number 0250080617
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/12/2018
Event Type  malfunction  
Manufacturer Narrative
The product was returned for investigation and the reported failure mode was confirmed.The failure mode will be monitored for future reoccurrence.Alleged failure: micro holes on shaft insulation.Confirmed failure: micro holes on shaft insulation.Probable root cause: poor autoclave reliability, incorrect sterilization/reprocessing procedure, handling procedures, contact forces, product used beyond defined useful life.(b)(4).
 
Event Description
It was reported that the insulation had been compromised.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PKG, 5MM PEEK MONOPOLAR HANDLE
Type of Device
ENDOSCOPIC GRASPING/CUTTING INSTRUMENT, NON-POWERED
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
roomi banerjee dua
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key7495132
MDR Text Key107703802
Report Number0002936485-2018-00408
Device Sequence Number1
Product Code OCZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K973259
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 05/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number0250080617
Device Lot Number1440288H  ¿
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/12/2018
Initial Date Manufacturer Received 04/12/2018
Initial Date FDA Received05/08/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/18/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-