| Catalog Number |
168-450 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
04/13/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported "fluid coming out" of device upon insertion and the device "was found with a tear in it." the device made contact with the patient.Surgery was completed with a back-up device.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: visual analysis of the returned device identified a broken shell, red particles, and flat creases.A leak test was performed and found a broken shell.A microscopic analysis identified a broken shell with striated edge on anterior side assessed as surgical damage.Fill inspection was performed and identified no blockage in the valve.Based on the device analysis the final assessment is: a broken shell with striated edge assessed as surgical damage.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported "fluid coming out" of device upon insertion and the device "was found with a tear in it." the device made contact with the patient.Surgery was completed with a back-up device.
|
| |
|