• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC MICROPUNCTURE TRANSITIONLESS ACCESS SET; DYB INTRODUCER, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC MICROPUNCTURE TRANSITIONLESS ACCESS SET; DYB INTRODUCER, CATHETER Back to Search Results
Catalog Number MPIS-401-NT-SST
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Pma/510(k) number: pre-amendment.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.
 
Event or Problem Description
It was reported that during a port-o-cath placement procedure, the micropuncture transitionless access set "broke" during cannulation through the skin.The complaint device was removed and another cannulation was performed.No adverse effects were reported.Additional information regarding event details, patient anatomy and outcome has been requested, but is not available at this time.
 
Additional Manufacturer Narrative
Investigation ¿ evaluation: a review of the dimensional verification, complaint history, device history record, documentation, quality control, specifications, and a visual inspection and functional evaluation of the returned device were conducted during the investigation.The inspection of the returned device confirmed that the inner dilator was bent approximately 50 degrees.The bend began approximately 2.5 cm from the proximal hub.The outer sheath was separated at a distance of 1.8 cm from the proximal hub, with stretching noted at the point of separation.The separated segment was measured to be 8.3 cm in length.Accordion-like damage was observed at the distal tip of the device.There were also surface marks on the shaft of the outer sheath.Additionally, a document based investigation evaluation was performed.There is no evidence to suggest the product was not made to specifications.A review of the device history record showed the lot was released meeting all finished goods release criteria.It should be noted there were no other reported complaints for this lot number.Based on the information provided, the examination of the returned product, and the results of our investigation, a definitive root cause could not be determined.Measures are being conducted to address this failure mode.We will continue to monitor for similar complaints.Per the risk assessment, no further action is required.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MICROPUNCTURE TRANSITIONLESS ACCESS SET
Common Device Name
DYB INTRODUCER, CATHETER
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key7496003
Report Number1820334-2018-01220
Device Sequence Number1339842
Product Code DYB
UDI-Device Identifier00827002479425
UDI-Public(01)00827002479425(17)200131(10)7638031
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberMPIS-401-NT-SST
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 04/17/2018
Supplement Date Received by Manufacturer06/20/2018
Initial Report FDA Received Date05/08/2018
Supplement Report FDA Received Date07/10/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-