| Model Number |
IMPELLA CP |
| Medical Device Problem Codes |
Kinked (1339); Difficult to Remove (1528); Device Issue (2379)
|
| Health Effect - Clinical Codes |
Perforation of Vessels (2135); Blood Loss (2597)
|
| Date of Event |
04/23/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The impella pump was returned for analysis.The data logs were not retrieved.Upon examination of the pump there was damage noted as the catheter was stretched and broken.Biomaterial was observed within the catheter's cannula.The pump housing was noted to be deformed and damaged.The root cause of the femoral artery dissection was most likely the excessive force used to remove the guidewire and pump.The root cause of the necessity for this excessive force is unable to be determined.No corrective action is warranted, the failure mode will be monitored and trended.Internal reference (b)(4).
|
| |
|
Event or Problem Description
|
|
A patient admitted with cardiogenic shock and acute mi had an impella cp placed with difficulty via the right femoral artery during cpr.Due to the pump kinking, when the guidewire was removed, the pump was explanted to resolve the kink.Upon reinsertion the right femoral artery was observed to be dissected.The patient had the dissection tamponade by the placement of a 14fr introducer and the case and cpr continued, until his death from cardiogenic shock and failed cpr and defibrillations.
|
| |
|