• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. FLUENCY PLUS ENDOVASCULAR STENT GRAFT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. FLUENCY PLUS ENDOVASCULAR STENT GRAFT Back to Search Results
Model Number FEM08120
Medical Device Problem Codes Positioning Failure (1158); Fracture (1260)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 03/19/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No medical records and no medical images were provided to the manufacturer.The lot number for the device was provided.The device history records are currently under review.The device has been returned for evaluation.The investigation is currently underway.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that during a lower left arm a/v graft procedure, the endovascular stent graft allegedly failed to deploy.Therefore, the catheter was removed without incident.There was no reported calcification nor tortuous path within the vessel.Reportedly, another endovascular stent graft was used to complete the procedure.There was no reported patient injury.
 
Additional Manufacturer Narrative
The lot number was provided and the lot device history records have been reviewed.The lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.This is the only complaint that has been reported for this corporate lot number to date.Investigation summary: based on the condition of the returned sample, the reported deployment failure was confirmed.The stent graft was found to be completely loaded and the outer sheath was found to be elongated indicating that increased friction must have affected the delivery system during the attempt to deploy the stent graft finally leading to the fracture of the outer sheath which made a stent graft deployment impossible.Therefore, as a result of the investigation performed the complaint was confirmed.However, based on the available information and the evaluation of the returned sample, a definite root cause for the reported event could not be determined.Labeling review: in reviewing the labeling supplied with this product it was found that the instructions for use (ifu) sufficiently address the potential risks.Regarding preparation of the device the ifu states: "prior to loading the endovascular system over a guide wire, both ports must be flushed with sterile saline.Flushing these lumens will also facilitate stent graft deployment." regarding the use of accessories the ifu states: "materials required for the fluency® plus endovascular stent graft procedure: introducer sheath with appropriate inner diameter".Furthermore, the ifu states: "if unusual resistance or high deployment force is encountered during stent graft deployment, abort the procedure, remove the delivery system and use an alternative device.".
 
Event or Problem Description
It was reported that during a lower left arm a/v graft procedure, the endovascular stent graft allegedly failed to deploy.Therefore, the catheter was removed without incident.There was no reported calcification nor tortuous path within the vessel.Reportedly, another endovascular stent graft was used to complete the procedure.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLUENCY PLUS ENDOVASCULAR STENT GRAFT
Common Device Name
ENDOVASCULAR STENT GRAFT
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key7496617
Report Number2020394-2018-00614
Device Sequence Number194737
Product Code PFV
UDI-Device Identifier04049519008523
UDI-Public(01)04049519008523
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P130029
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2020
Device Model NumberFEM08120
Device Catalogue NumberFEM08120
Device Lot NumberANBW1007
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/11/2018
Initial Date Received by Manufacturer 04/11/2018
Supplement Date Received by Manufacturer06/03/2018
Initial Report FDA Received Date05/08/2018
Supplement Report FDA Received Date06/25/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age82 YR
Patient Weight54
-
-