| Catalog Number |
369741 |
| Medical Device Problem Code |
Short Fill (1575)
|
| Health Effect - Clinical Code |
No Information (3190)
|
| Date of Event |
02/05/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
"multiple lot numbers: there were multiple lot numbers reported to be involved.The information for the additional lot number is as follows: medical device lot #: 7100570.Medical device expiration date: 10/31/2018.Returned to manufacturer on: unknown.Device manufacture date: 04/10/2017." bd received samples from the customer facility for investigation.The samples were evaluated and the customer's indicated failure mode with the incident lot was not observed.A review of the manufacturing records was completed for the incident lot and no issues were identified.Bd has initiated a capa to document further investigation and root cause(s) of this product issue.
|
| |
|
Event or Problem Description
|
|
It was reported that the bd vacutainer® esr blood collection tubes under filled.
|
| |
|