• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: JOHNSON & JOHNSON SURGICAL VISION, INC. TECNIS 1 MULTIFOCAL; MULTIFOCAL IOLS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

JOHNSON & JOHNSON SURGICAL VISION, INC. TECNIS 1 MULTIFOCAL; MULTIFOCAL IOLS Back to Search Results
Model Number ZLB00
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 10/11/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Unknown, information not provided.If explanted, give date: not applicable, as the lens remains implanted.  lens repositioned.Dysphotopsia - negative.It was indicated that the device is not returning for evaluation; therefore a failure analysis of the complaint device cannot be completed.A review of the device/lot history record, labeling, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that tecnis multifocal intraocular lens (model zlb00, +25.5 diopter) was repositioned post operative because of patient experiencing negative dysphotopsia (temporal shadow) post op in the left eye (os) that would not subside.The doctor repositioned the lens with the haptics oriented horizontally to mitigate the effects of temporal shadowing.The lens was not explanted.No enlargement of the incision occurred.The repositioning was performed without incident.The patient was reported to have no complaints post repositioning.No further information provided.
 
Additional Manufacturer Narrative
Additional information was received.It was reported the patient was not happy with the lens from the start.The lens was repositioned on (b)(6) 2018 and the patient was still not happy.It was learned that the explant was completed on (b)(6) 2018.Reportedly lens from another manufacturer was used as replacement lens for the patient gender/sex: male.Age: 66.If explanted, give date: (b)(6) 2018.All pertinent information available johnson & johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
Device available for evaluation: yes.Returned to manufacturer on: 8/1/2018.Device returned to manufacturer: yes.Device evaluation: the product was returned to the manufacturing site for evaluation.The sample was inspected by a qualified final inspection operator using 12x magnification.There are no anomalies found.The complaint cannot be confirmed.Manufacturing records review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.A search on complaints revealed that no similar complaints were received for this production order number.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions along with warnings for the proper use and handling of the device.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS 1 MULTIFOCAL
Common Device Name
MULTIFOCAL IOLS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
santa ana CA
MDR Report Key7497396
Report Number9614546-2018-00445
Device Sequence Number13872503
Product Code MFK
UDI-Device Identifier05050474560789
UDI-Public(01)05050474560789(17)191005
Combination Product (Y/N)Y
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/21/2018
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Physician
Device Expiration Date10/05/2019
Device Model NumberZLB00
Device Catalogue NumberZLB00U0255
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/01/2018
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 04/10/2018
Supplement Date Received by Manufacturer05/17/2018
09/13/2018
Initial Report FDA Received Date05/08/2018
Supplement Report FDA Received Date06/12/2018
09/21/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age66 YR
Date Report Sent to FDA01/01/2005
-
-