| Catalog Number |
TH-LX-M |
| Medical Device Problem Code |
Device Operational Issue (2914)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
04/10/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Evaluation summary: two turbohawk devices were returned one of the turbohawk units was connected to a cutter driver.No other ancillary units were included.The turbohawk connected to the cutter driver was inspected.The cutter blank was visible in the cutter window.The edge of the cutter was approximately 0.3cm distal the cutter window.Traces of dried blood/biological debris was observed at the cutter mouth and within the cutter window.The thumb-switch was retracted and the cutter pulled back into the cutter.No damages were noted.The thumb-switch was advanced with the cutter driver activated.The cutter advanced approximately 0.8cm distal the cutter window.It should be noted the cutter stopped at the location of the plunger.A 0.014 guidewire from the lab was inserted the cutter mouth of the distal assembly and pushed the plunger towards the cutter window after the cutter was retracted back into the cutter window.Resistance was first encountered while attempting to push the plunger.The plunger then was able to be pushed back into the cutter window.The thumb-switch was advanced a second time.The cutter was able to advance approximately 2.9cm distal the cutter window.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Physician intended to use a turbohawk device for treatment of a slightly calcified soft tissue lesion with little tortuosity in the patients mid superficial femoral artery (sfa).Ifu was followed and the device was prepped and inspected prior to use without issue.It is reported the device became jammed and would not close.The physician could turn the cutter off but not completely forward.The cutter was barely closed when the device was removed from the patient.The same issue is reported for a second turbohawk device used in the procedure.A third device h1-lx was used to complete the procedure.No patient injury reported.
|
| |
|
Search Alerts/Recalls
|