| Model Number |
956-616 |
| Medical Device Problem Codes |
Fluid/Blood Leak (1250); Fracture (1260)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
01/01/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Date of event is best guess.
|
| |
|
Event or Problem Description
|
|
According to the complaint, blood sprayed uncontrollably out from the syringe.The user (a nurse) was not in contact with blood.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation has identified that the root cause is a manufacturing error related to the labelling assembly in the production.
|
| |
|
Additional Manufacturer Narrative
|
|
Besides the fact that the investigation has identified that the root cause is a manufacturing error related to the labelling assembly in the production, the implemented production changes are: extension of 100% visual inspection to cover all products on the affected production line.Changes on the labeling station: rotation angle: rotation of cylinder during labelling process.Velocity: speed of cylinder's rotation during labelling process.Dispenser speed: speed of feeding label during labelling process.
|
| |
|