| Catalog Number |
050-87222 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
04/20/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The plant investigation is in progress.A supplemental medwatch report will be submitted upon completion of this activity.
|
| |
|
Event or Problem Description
|
|
It was reported that a fluid leak was identified coincident with peritoneal dialysis therapy.The patient stated that during the first fill the cassette leaked fluid from the pin.The patient confirmed that no fluid was on the cycler.The patient was advised to cancel treatment and setup with new supplies to complete treatment.During follow up, the pdrn confirmed the patient did not develop any symptoms, adverse events, injuries, or require medical intervention as a result of the reported event.The pdrn stated that the patient was able to continue pd therapy with no further issues.
|
| |
|
Additional Manufacturer Narrative
|
|
Plant investigation: as the device was not returned to the manufacturer, a physical evaluation could not be performed.A batch records review was conducted by the manufacturer for the reported lot.There were no non-conformances or abnormalities identified during the manufacturing process which could be associated with the reported event.The entire lot has been sold and distributed.In addition, a device history review was performed and confirmed that the results of the in-progress and final quality control (qc) testing met all requirements.The lot met all specifications for release.A product history review did not reveal a probable cause for the customer complaint.As a physical evaluation could not be performed, a definitive conclusion regarding the reported incident could not be reached and a cause could not be confirmed.
|
| |
|
Search Alerts/Recalls
|