ST. JUDE MEDICAL - NEUROMODULATION PROCLAIM¿ 7 IPG; SCS IPG Back to Search Results
Model Number 3667
Device Problem Device Inoperable (1663)
Patient Problem Inadequate Pain Relief (2388)
Event Date 04/10/2018
Event Type  Injury  
Manufacturer Narrative
The device is included in the neuromodulation implantable pulse generator (ipg) inoperable when exposed to monopolar electrosurgery advisory notice issued by abbott on 02 june 2017.The manufacturer has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.The manufacturer defers to the patient's physician regarding medical history.
 
Event Description
It was reported (belgium) the patient's ipg became inoperable after undergoing hip surgery.Surgical intervention is planned to address the issue.
 
Event Description
Additional information identifed that surgical intervention was undertaken wherein the ipg was explanted and replaced.Therapy was restored post-operatively.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROCLAIM¿ 7 IPG
Type of Device
SCS IPG
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
MDR Report Key7498456
MDR Text Key107726196
Report Number1627487-2018-04279
Device Sequence Number1
Product Code LGW
Combination Product (y/n)N
Reporter Country CodeBE
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial,Followup,Followup
Report Date 07/16/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/09/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/05/2019
Device Model Number3667
Device Lot Number6046526
Other Device ID Number05415067023254
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/06/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received07/10/2018
Was Device Evaluated by Manufacturer? No
Date Device Manufactured07/05/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction Number1627487-060217-001-C
Patient Sequence Number1
Patient Outcome(s) Other;
-
-