• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. LIGACLIP*SINGLE CLIP APPLIER; LIGACLIP APPLIER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. LIGACLIP*SINGLE CLIP APPLIER; LIGACLIP APPLIER Back to Search Results
Catalog Number LC310
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Only event year known: 2018.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
 
Event or Problem Description
It was reported that during an unknown procedure, clip malformation.There were no patient consequences.
 
Additional Manufacturer Narrative
(b)(4).Device analysis: the analysis results found that the lc310 device was received with no apparent damage.In an attempt to replicate the reported incident, the device was tested for functionality.Upon functional testing of the device, the instrument loaded, retained and deployed 6 clips as intended.The instrument was fully functional and conforming to our manufacturing requirements.No conclusion could be reached as to what may have caused the reported incident.The batch history records were reviewed and certed by external manufacturing that the manufacturing criteria was met prior to the release of the equipment. .
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
LIGACLIP APPLIER
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
MDR Report Key7498493
Report Number3005075853-2018-09711
Device Sequence Number167727
Product Code GDO
UDI-Device Identifier20705036012662
UDI-Public20705036012662
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberLC310
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/11/2018
Initial Date Received by Manufacturer 04/18/2018
Supplement Date Received by Manufacturer05/11/2018
Initial Report FDA Received Date05/09/2018
Supplement Report FDA Received Date05/31/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-