| Model Number |
2192 |
| Medical Device Problem Code |
Difficult to Remove (1528)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, the balloon was stuck inside the patient's body.There was no reported patient involvement and outcome.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation technical support confirmed the calibration error.A visual inspection showed no issues there was no external damage to the probe.The unit was tested and showed calibration errors on channel 6 and b1.There were no visible signs of mishandling/damage.The investigation found the most probable cause to be the sensor failure.But a root cause was not identified.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, the balloon was stuck inside the patient's body.The device has been returned for investigation.The unit has been tested and showed calibration errors on channel 6 and 31 with no visible signs of mishandling or damage.There was no reported patient involvement and outcome.
|
| |
|