| Catalog Number |
9001C0212 |
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
04/18/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The device has not been returned, at this time, to the manufacturer for evaluation.A lot history review (lhr) of rebt2285 showed no other similar product complaint(s) from this lot number.
|
| |
|
Event or Problem Description
|
|
It was reported that the device presented a hole.It was further reported that the cover presented a hole as well.The device was not used on a patient.
|
| |
|
Additional Manufacturer Narrative
|
|
The following were reviewed as part of this investigation: patient severity, frequency analysis, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the complaint of a hole present in the probe cover is confirmed and appears to be use related.One site rite needle guide kit packaging with lot: rebt2285 and ref: 9001c0212 and one probe cover was returned for investigation.The probe cover was covered in gel.The probe cover was filled with water to determine the hole location.Two leaks were observed approximately 1.3 inches from the closed end of the cover.The leak locations were approximately 1 inch apart.A scratch in the cover was observed near the two leak locations.Microscopic observation revealed a slit in the plastic cover at the leak sites.The scratch in the cover and the two slits appear to have been caused by contact with a sharp instrument; therefore, the complaint is confirmed, use related.A lot history review (lhr) of rebt2285 showed no other similar product complaint(s) from this lot number.
|
| |
|
Event or Problem Description
|
|
It was reported that the device presented a hole.It was further reported that the cover presented a hole as well.The device was not used on a patient.
|
| |
|
Search Alerts/Recalls
|