• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS, INC. 1818910 ASPHERE M SPEC 12/14 40 +8.5; HIP METAL FEMORAL HEADS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS, INC. 1818910 ASPHERE M SPEC 12/14 40 +8.5; HIP METAL FEMORAL HEADS Back to Search Results
Catalog Number 136507100
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Unspecified Infection (1930); Pain (1994); Weakness (2145); Distress (2329); Injury (2348); Limited Mobility Of The Implanted Joint (2671); Not Applicable (3189); No Code Available (3191)
Date of Event 11/26/2016
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Investigation summary: no device associated with this report was received for examination.A worldwide complaint database search found no other reported incident(s) against the provided product/lot combination(s) since release for distribution.Based on previous investigations this complication of joint replacement is unlikely to have been the result of a device failing to meet required specifications.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Litigation received.Litigation alleges injury, pain, elevated metal ions, walking difficulty, weakness, lack of mobility and emotional distress.It was also stated in litigation that during procedure the surgeon found a large amount of purulent fluid which was infected.Head and liner were removed and implanted with depuy poly liner and ceramic head.Laboratory results for metal ions were below 7(no unit provided).Doi: (b)(6) 2011; dor: (b)(6) 2016 ; left hip.(b)(4)- 2nd revision.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report. udi:(b)(4).
 
Event or Problem Description
Ppf alleges elevated metal ions, though it was mentioned that there is elevated metal ions but laboratory results were below 7.Added stem for the alleged elevated metal ions and patient age.Doi: (b)(6) 2011.Dor: (b)(6) 2016.Left hip.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provision of 21 cfr, part 803.The report may be based on the information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).No device associated with this report was received for examination.A worldwide complaint database search found no other reported incident(s) against the provided product/lot combination(s) since release for distribution.Based on previous investigations this complication of joint replacement is unlikely to have been the result of a device failing to meet required specifications.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Device history lot : null.Device history batch : null.Device history review : null.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASPHERE M SPEC 12/14 40 +8.5
Common Device Name
HIP METAL FEMORAL HEADS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic drive
warsaw IN 46582 0988
MDR Report Key7499396
Report Number1818910-2018-59224
Device Sequence Number19320134
Product Code JDI
UDI-Device Identifier10603295033011
UDI-Public10603295033011
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K082585
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2011
Device Explanted Year2016
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/23/2015
Device Catalogue Number136507100
Device Lot Number3201673
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 04/12/2018
Supplement Date Received by Manufacturer03/08/2019
07/11/2019
Initial Report FDA Received Date05/09/2018
Supplement Report FDA Received Date03/22/2019
07/12/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age69 YR
-
-