Model Number H7493926028220
Medical Device Problem Codes Bent (1059); Device Damaged by Another Device (2915)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 04/11/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
Same case as mdr id: 2134265-2018-03970.It was reported that stent damage occurred.The target lesion was located in a severely calcified vessel.After an 8f guidezilla¿ ii guide extension catheter was positioned, a 2.25 x 28 synergy ii drug-eluting stent was advanced to treat the lesion.However, as the stent was being pulled back into the guidezilla, it was noted that the guidezilla crunched up the stent.Both ends of the stent struts were shortened.The procedure was completed using a different device.No patient complications were reported and the patient's status was fine.
 
Additional Manufacturer Narrative
Device evaluated by mfr., eval summary attached, method codes, result codes, conclusion codes updated.Device evaluated by mfr.: stent delivery system was returned for analysis.A visual examination of the stent found that distal stent rows 1-3 were damaged with stent struts opened and lifted from their crimped stent position.The undamaged crimped stent od (outer diameter) was measured and the result was within maximum crimped stent profile measurement.The balloon cones were reviewed and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and microscopic examination of the bumper tip showed signs of damage.A visual and tactile examination of the hypotube found no issues.A visual and tactile examination of the outer and inner lumen and mid-shaft section found no issues along the shaft polymer extrusion.No other issues were identified during the product analysis.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The investigation conclusion is caused by other device as another device/drug/subsequent procedure caused the complaint event.(b)(4).
 
Event or Problem Description
Same case as mdr id: 2134265-2018-03970.It was reported that stent damage occurred.The target lesion was located in a severely calcified vessel.After an 8f guidezilla¿ ii guide extension catheter was positioned, a 2.25 x 28 synergy ii drug-eluting stent was advanced to treat the lesion.However, as the stent was being pulled back into the guidezilla, it was noted that the guidezilla crunched up the stent.Both ends of the stent struts were shortened.The procedure was completed using a different device.No patient complications were reported and the patient's status was fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNERGY II EVEROLIMUS-ELUTING PLATINUM CHROMIUM CORONARY STENT SYSTEM
Common Device Name
BIODEGRADABLE POLYMER DRUG ELUTING CORONARY STENT SYSTEM
Manufacturer (Section D)
BOSTON SCIENTIFIC - GALWAY
MDR Report Key7499556
Report Number2134265-2018-03969
Device Sequence Number13947230
Product Code NIQ
UDI-Device Identifier08714729840268
UDI-Public08714729840268
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
P150003
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/16/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/06/2018
Device Model NumberH7493926028220
Device Catalogue Number39260-2822
Device Lot Number0021515646
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/03/2018
Initial Date Received by Manufacturer 04/16/2018
Supplement Date Received by Manufacturer05/24/2018
Initial Report FDA Received Date05/09/2018
Supplement Report FDA Received Date06/13/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
GUIDEZILLA¿ II GUIDE EXTENSION CATHETER
-
-