Model Number BL+A223D/V809D
Medical Device Problem Codes Obstruction of Flow (2423); Device Operational Issue (2914)
Health Effect - Clinical Codes Syncope (1610); Diaphoresis (2452)
Date of Event 03/30/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Investigation report attached is on retained samples only, pending return of actual used device from customer.
 
Event or Problem Description
On (b)(6) 2018, baxter product surveillance received a call from an administrator regarding an arterial line that was filled with air (product code: (b)(4), lot #: 17k-20-6 or 17k-28-9).The administrator stated that there was no drug associated with the treatment.The administrator stated that the patient became hypotensive, diaphoretic and lost consciousness.They gave oxygen support since the patient had an agonal breathing and they bagged the patient and the patient regained consciousness.The patient was placed in a left side lying position and was sent to the hospital.They performed a cat scan and saw that there was no air in the system.They kept the patient in the hospital overnight for observation.The administrator stated that they don't know the exact lot number which had the air in the arterial line.The sample is available.There was patient involvement and patient injury was indicated at the time of initial report.
 
Event or Problem Description
On 10-apr-2018, baxter product surveillance received a call from an administrator regarding an arterial line that was filled with air (product code: a223dv809d, lot #: 17k-20-6 or 17k-28-9).The administrator stated that there was no drug associated with the treatment.The administrator stated that the patient became hypotensive, diaphoretic and lost consciousness.They gave oxygen support since the patient had an agonal breathing and they bagged the patient and the patient regained consciousness.The patient was placed in a left side lying position and was sent to the hospital.They performed a cat scan and saw that there was no air in the system.They kept the patient in the hospital overnight for observation.The administrator stated that they don't know the exact lot number which had the air in the arterial line.The sample is available.There was patient involvement and patient injury was indicated at the time of initial report.
 
Additional Manufacturer Narrative
Investigation report attached is on retained samples only, pending return of actual used device from customer.8/30/18: device was never returned by the customer for factory evaluation.Final investigation attached on retained samples only.Revised section h6: revised & added event problem and evaluation codes: device code from 2914 to 2423.Method code removed codes from 38, 3317, 3330, 3340 to 11, 3331, 4114.Conclusion code from 92 to 4310 & 67.8/30/18: removed conclusion code 92.
 
Additional Manufacturer Narrative
Investigation report attached is on retained samples only, pending return of actual used devce from customer.8/30/18: device was never returned by the customer for factory evaluation.Final investigation attached on retained samples only.Revised section e3, g4, and g7 revised section h6: revised & added event problem and evaluation codes: device code from 2914 to 2423 method code removed codes from 38, 3317, 3330, 3340 to 11, 3331, 4114 concluclusion code from 92 to 4310 & 67.
 
Event or Problem Description
On 10-apr-2018, baxter product surveillance received a call from an administrator regarding an arterial line that was filled with air (product code: a223dv809d., lot #: 17k-20-6 or 17k-28-9).The administrator stated that there was no drug associated with the treatment.The administrator stated that the patient became hypotensive, diaphoretic and lost consciousness.They gave oxygen support since the patient had an agonal breathing and they bagged the patient and the patient regained consciousness.The patient was placed in a left side lying position and was sent to the hospital.They performed a cat scan and saw that there was no air in the system.They kept the patient in the hospital overnight for observation.The administrator stated that they don't know the exact lot number which had the air in the arterial line.The sample is available.There was patient involvement and patient injury was indicated at the time of initial report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NIPRO BLOOD TUBING SET W/TRANSDUCER PROTECTOR
Common Device Name
BLOODLINE
Manufacturer (Section D)
NIPRO (THAILAND) CORP., LTD.
10/2 moo 8,
bangnomko, sena
ayuthaya, thailand, 13110
TH  13110
Manufacturer (Section G)
NIPRO (THAILAND) CORP., LTD.
10/2 moo 8,
bangnomko, sena
ayuthaya, thailand, 13110
TH   13110
Manufacturer Contact
michelle tejada
3150 nw 107 avenue
miami, FL 33172
3055997174
MDR Report Key7499625
Report Number8041145-2018-00016
Device Sequence Number1963455
Product Code FJK
UDI-Device Identifier00383790004176
UDI-Public00383790004176
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K072024
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/30/2018
Report Date (Section F) 05/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberBL+A223D/V809D
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date04/16/2018
Event Location Outpatient Treatment Facility
Date Report to Manufacturer05/09/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer04/16/2018
Not provided
Initial Report FDA Received Date05/09/2018
Supplement Report FDA Received Date09/06/2023
08/30/2018
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
Patient SexUnknown
Date Report Sent to FDA05/09/2018
-
-