| Model Number |
407652 |
| Medical Device Problem Codes |
Failure to Capture (1081); Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
04/12/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that post-operatively, the right atrial (ra) lead became dislodged and was confirmed by x-ray to be in the right ventricle.No capture was exhibited in the atrium.The ra lead was explanted and replaced.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Product event summary: the full lead was returned, analyzed, and no anomalies were found.The proximal conductor of the lead became extrinsically distorted due to pulling/stretching/overstress.The distal low voltage electrode of the lead was covered in body tissue/fibrotic growth.The outer insulation of the lead was extrinsically breached due to a cut.Visual analysis of the lead indicated damage at implant.If information is provided in the future, a supplemental report will be issued.
|
| |
|