(b)(4).Device was used for treatment, not diagnosis.The complaint device was received and evaluated.Visual observation reveals that the anchor positioned at the tip of the inserter is not aligned/bent to the inserter shaft.There were stains near the inserted distal end and even the suture is discolored near the anchor confirming that this is not an out of the box failure and the device has been used.So, the reported failure cannot be confirmed.To investigate further, the anchor was removed from the tip and it was noticed that the inserter tip was bent which caused the anchor to look bent.Inserter tip bent might be caused due to off-axis insertion, levering during insertion or hard bone quality.A dhr review has been conducted and the results indicate that this batch of product was processed without incident and therefore there is no internally assignable cause for the reported problem.Further, a review into the depuy synthes mitek complaints system revealed no other complaint for this lot of devices that were released to distribution.At this point in time, no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Udi: (b)(4).
|