• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC STERNAL ZIPFIX WITH NEEDLE STERILE/20 PACK; CERCLAGE FIXATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC STERNAL ZIPFIX WITH NEEDLE STERILE/20 PACK; CERCLAGE FIXATION Back to Search Results
Model Number 08.501.001.20S
Medical Device Problem Code Device Operates Differently Than Expected (2913)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 04/11/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Patient information is not available for reporting.Due to the intra-operative events, the device was not successfully implanted.An alternate device was used to complete procedural step.As such, implant/explant dates are not applicable.Complainant part is not expected to be returned for manufacturer review/investigation.A device history record (dhr) review was performed for part no.: 08.501.001.20s, lot no.: l656550: manufacturing location: (b)(4), supplier: (b)(4), release to warehouse date: 06.Dec.2017, expiry date: 01.Nov.2022.No non-conformance reports (ncrs) were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of the product that would contribute to this complaint condition.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Device report from synthes an event in (b)(6) as follows: it was reported that during an unknown procedure on (b)(6) 2018, the surgeon put the patient safety & learning system (psls) for zipfix in patient and secured it but it came undone.The surgeon was unable to re-insert it.A new zipfix strap was opened and the case was completed.No additional information is available.This report is for sternal zipfix with needle sterile.This is report 1 of 1 for complaint (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STERNAL ZIPFIX WITH NEEDLE STERILE/20 PACK
Common Device Name
CERCLAGE FIXATION
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key7500239
Report Number2939274-2018-52120
Device Sequence Number11250072
Product Code JDQ
UDI-Device Identifier20887587025992
UDI-Public(01)20887587025992
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K110789
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Health Professional
Type of Report Initial
Report Date (Section B) 04/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/01/2022
Device Model Number08.501.001.20S
Device Catalogue Number08.501.001.20S
Device Lot NumberL656550
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 04/17/2018
Initial Report FDA Received Date05/09/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-