Model Number 1884002
Medical Device Problem Code Material Fragmentation (1261)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 03/26/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product analysis confirmed that one device, part number 1884002 from lot hg1wygp, was received with the inner cutter tip broken off.A microscope and calipers were used to further assess the returned sample.The fragmented portion was approximately 0.24" in length with the detachment site occurring at the first proximal tooth valley.The inner assembly was deformed in such a way that indicates the inner blade teeth impacted the outer cutting surface while moving both the clockwise and counterclockwise directions which resulted in the observed break.The impacts markings were at the second and third proximal valleys, and there was uneven wear at the cutting tip which may indicate an interference fit between the inner and outer assemblies at the tip.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
The customer reported that the tip of the blade broke off.The tip was retrieved.There was a delay of less than 10 minutes and no patient injury or impact.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XPS® BLADE
Common Device Name
BUR, EAR, NOSE AND THROAT
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
MDR Report Key7500276
Report Number3004209178-2018-10494
Device Sequence Number144474
Product Code EQJ
UDI-Device Identifier20681490047443
UDI-Public20681490047443
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/22/2025
Device Model Number1884002
Device Catalogue Number1884002
Device Lot NumberHG1WYGP
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/03/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/12/2018
Initial Report FDA Received Date05/09/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-