| Model Number |
E SERIES |
| Medical Device Problem Code |
Device Displays Incorrect Message (2591)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during functional testing, the device displayed a "defib fault 78" message.Complainant did not indicate that there was any patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to zoll medical corporation; the malfunction was duplicated and attributed to a system interconnect flex cable not seated properly.A system interconnect flex cable and connector interlock were replaced to remedy the problem.Analysis for reports of this type has not identified an increase in trend.
|
| |
|