| Model Number |
H7493952824300 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
04/13/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device is a combination product.(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that shaft break occurred.The target lesion was in the left coronary vessel.A 3.00x24mm promus premier¿ stent was advanced to treat the lesion.However, the device failed to cross the lesion and it was noted that the shaft broke.The procedure was completed with another of the same device.No patient complications were reported and the patient's status was fine.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.: the stent delivery system was returned for analysis.A visual and microscopic examination of the stent found no issues.There was no sign of damage, stretching or lifting of the stent struts.The stent showed no signs of movement and was set equidistant between the proximal and distal markerbands.The crimped stent od (outer diameter) was measured within maximum crimped stent profile measurement.A visual and microscopic examination of the tip found damage.The balloon cones were reviewed and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and tactile examination of the device found a hypotube break 570 mm distal from the distal end of the strain relief.Multiple hypotube kinks were also noted along the full length of the catheter.This type of damage is consistent with excessive force being applied to the delivery system.A visual and tactile examination of the inner, outer and midshaft section found no issues along the shaft polymer extrusion.No other issues were identified during the product analysis.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The investigation conclusion is operational context as the product meets the design and manufacture specification but due to anatomical/procedural factors encountered during the procedure, performance was limited.(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that shaft break occurred.The target lesion was in the left coronary vessel.A 3.00x24mm promus premier stent was advanced to treat the lesion.However, the device failed to cross the lesion and it was noted that the shaft broke.The procedure was completed with another of the same device.No patient complications were reported and the patient's status was fine.
|
| |
|