Catalog Number XR30V
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Batch # n5788y.Event month and day unknown.Only year (2017) is known.Device evaluation: the analysis results showed that the xr30v reload arrived with no apparent damaged and partially loaded with only 12 staples present and all protruding.No functional test could be performed due to the conditions of the reload.It should be noted that when manipulating the device only the closing trigger should be grasped until ready to fire the device.Do not grasp the firing trigger before the device is to be fired.If the firing trigger is manipulated it could cause the staples to deploy partially or completely resulting in malformed staples and a premature lockout situation.For further details please refer to the instruction for use.The batch history record was reviewed and no defects, ncr¿s or protocols related to the complaint, were found during the manufacturing process.
 
Event or Problem Description
It was reported that during an unknown procedure, the firing trigger wasn't available, even though tissue wasn't that thick (it was a vessel).The surgeon changed to another cartridge, and it worked well.There were no patient consequences reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
MDR Report Key7500580
Report Number3005075853-2018-09732
Device Sequence Number1813757
Product Code GDW
UDI-Device Identifier20705036013454
UDI-Public20705036013454
Combination Product (Y/N)N
PMA/510(K) Number
K020779
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 12/17/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/18/2021
Device Catalogue NumberXR30V
Device Lot NumberN4MA96
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/19/2018
Initial Date Received by Manufacturer 04/19/2018
Initial Report FDA Received Date05/09/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-