| Catalog Number |
X558H |
| Medical Device Problem Code |
Tear, Rip or Hole in Device Packaging (2385)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device history records were reviewed and the manufacturing criteria were met prior to the release of this lot.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Additional information was requested and the following was obtained: how many suture packages were noticed to have tears in the sterile tyvek? - unknown.What are the product code(s) and lot number(s) of the sutures with tears in the sterile tyvek? -unknown.What were the storage conditions? - facility claims stored in sterile core within pyxis machine, removed from box and placed into plastic containers.Were any anomalies noted at the time of initial receipt of this ethicon product? no.Were they reported to ethicon? na.What was the condition upon initial removal from the original sealed ethicon boxes prior to entry into pyxis machine? nothing reported.What are the environmental and temperature conditions in storage and the pyxis machine? per logistics, no alarms¿however sales rep discussed hurricane harvey/aug.2017, likely power outages.Is this a new pyxis system at the facility? in use for some time.Sales rep physically has all product mentioned.Suggested temp readings / mention hurricane harvey-power outages.Distributed by (b)(4) /went to them w/complaint 1st.No practices / change.These products are close to exp.Dates ¿ have been in possession for some time.Fine when placed and removed from boxes.Nothing noted.Nothing crumbled when placed in caddy by logistics person ((b)(6)-no alarms-indicating variance in temperature on pyxis) , not from one order, received over time into inventory.No specific information regarding quantity of devices has been provided.If further information is received, the file will be reviewed and updated accordingly.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent a surgical procedure and suture was used.While attempting to open a suture pack into sterile field, nurse felt inner suture retention plastic crumble within her fingers.Suture not introduced into field, another pulled.Hospital logistics notified and began checking plastic tray for all suture packs.Pulled off the shelf all packages where tray crumbled to the touch, some packages had tears in the sterile tyvek warranting the suture non-sterile.There were no adverse patient consequences reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # pc-000171629 date sent to the fda: 07/11/2018.Evaluation unopened samples of product were returned for analysis.During the visual inspection of unopened samples, delamination and broken tyvek was noted on three packets, the rest were as expected.The samples were opened and an unusual color at the paper lids and the winding formers broken were observed.The manufacturing records were reviewed and the manufacturing/packaging criteria were met prior to the release of this batch.Per the sample condition the assignable cause packaging damaged - primary package and packaging integrity is caused by an external cause.
|
| |
|
Search Alerts/Recalls
|