| Model Number |
CQ7588 |
| Medical Device Problem Codes |
Peeled/Delaminated (1454); Unraveled Material (1664)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
02/07/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
No medical records or no medical images have been made available to the manufacturer.As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
|
| |
|
Event or Problem Description
|
|
It was reported that post removal of a pta balloon dilatation catheter, unraveled material was alleged found.There was no reported patient injury.
|
| |
|
Event or Problem Description
|
|
It was reported that post removal of a pta balloon dilatation catheter, unraveled material was alleged found.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Labeling review: the review of the ifu (instructions for use), indications, warnings, precautions, cautions, possible complications, and contraindications showed that the product labeling is adequate.Investigation summary: the device was returned for evaluation.A visual inspection found the pebax to be peeled from the balloon throughout the length of the balloon.A portion of the pebax was returned separate from the device.No unraveled fibers or fiber disturbance was noted to the balloon.Therefore, the investigation is confirmed for peeled pebax.The definitive root cause for the identified peeled pebax could not be determined based upon the available information.
|
| |
|