Catalog Number ATG80124
Medical Device Problem Codes Break (1069); Retraction Problem (1536); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 04/12/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No medical records or no medical images have been made available to the manufacturer.As the lot number for the device was provided, a review of the device history records is currently being performed.The device has not been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that during pta balloon dilatation, the balloon component detached off the catheter.The patient status is unknown.
 
Event or Problem Description
It was reported that during pta balloon dilatation, the balloon component detached off the catheter.The patient status is unknown.
 
Additional Manufacturer Narrative
Manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.This is the only complaint reported to date for this lot number and failure mode.Labeling review: the review of the ifu (instructions for use), indications, warnings, precautions, cautions, possible complications, and contraindications showed that the product labeling is adequate.Investigation summary: the device was returned for evaluation.A visual inspection found the balloon returned in the user's introducer sheath.A complete break of the proximal balloon joint was identified; the inner lumen was still intact.Further evaluation of the device found the inner lumen to be stretched indicating retraction issues.Additionally, the distal end of the sheath was flared.Therefore, the investigation is confirmed for a break at the proximal balloon joint, and for sheath-related retraction issues.The investigation is unconfirmed for the reported detachment, as the balloon was still attached to the device.It is likely that the user perceived the break as a detachment.Based on the condition of the returned sample, it is likely that resistance was encountered by the user leading to excessive force being applied to the device as it was being withdrawn through the sheath.This excessive force led to the balloon breaking from the shaft.However, the definitive root cause for the identified retraction issues could not be identified based on the available information.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATLAS GOLD PTA BALLOON DILATATION CATHETER
Common Device Name
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key7501214
Report Number2020394-2018-00636
Device Sequence Number1768420
Product Code LIT
UDI-Device Identifier00801741060694
UDI-Public(01)00801741060694
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K122984
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/09/2021
Device Catalogue NumberATG80124
Device Lot Number93PC0007
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/04/2018
Initial Date Received by Manufacturer 04/13/2018
Supplement Date Received by Manufacturer07/31/2018
Initial Report FDA Received Date05/09/2018
Supplement Report FDA Received Date08/15/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-