• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HALYARD HEALTH INTRODUCER KIT FOR GASTROSTOMY FEEDING TUBE; DH EF PERC PLACEMENT PRODUCTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HALYARD HEALTH INTRODUCER KIT FOR GASTROSTOMY FEEDING TUBE; DH EF PERC PLACEMENT PRODUCTS Back to Search Results
Model Number 98434-02
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 04/13/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product involved in the report has been returned and is being processed for evaluation.A review of the device history record is in-progress.Upon completion of the sample evaluation and investigation; a follow-up report will be filed.All information reasonably known as of (b)(6) 2018 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by halyard health represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to halyard health.Halyard health has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the halyard health complaint database and identified as complaint (b)(4).
 
Event or Problem Description
It was reported that 2 t-fasteners broke after placement.Only 2 of the 4 fasteners were used, and both broke after the dilator was in place.The physician ended up placing the feeding tube without the use of the t-fasteners.There was no patient injury or intervention required.
 
Additional Manufacturer Narrative
Halyard health received two (2) sutures (gastropexy needle) that were returned without the original case packaging.They were returned in the gastropexy tray with no product label/information returned.The outer needle shield exhibited slight staining (brownish) color.The two returned sutures, which were identified by the customer/user as being unused, were still decontaminated due to the fact that they were received inside a biohazard bag.The complaint description also identified that the two unused sutures would be returned as representative samples.After decontamination, the two returned sutures were inspected and did not exhibit any suture break.The appearance of the returned unused sutures noticed to be damaged free.The broken sutures that were mentioned in the complaint description were not returned to the manufacturer.The root cause of the reported issue could not be conclusively determined since the broken sutures that were mentioned in the complaint description were not returned to the manufacturer, only two unused sutures were received.The device history record for the reported lot number was reviewed and documented that the lot was manufactured according to the approved manufacturing procedures, specifications, and then released by quality assurance.All information reasonably known as of 27-jun-2018 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by halyard health represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to halyard health.Halyard health has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the halyard health complaint database and identified as complaint (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTRODUCER KIT FOR GASTROSTOMY FEEDING TUBE
Common Device Name
DH EF PERC PLACEMENT PRODUCTS
Manufacturer (Section D)
HALYARD HEALTH
5405 windward parkway
alpharetta GA 30004
MDR Report Key7501862
Report Number9611594-2018-00066
Device Sequence Number13501206
Product Code KGC
UDI-Device Identifier10680651984347
UDI-Public10680651984347
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K080253
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/29/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date08/01/2019
Device Model Number98434-02
Device Catalogue Number991098434
Device Lot NumberAA7254R06
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/25/2018
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 04/13/2018
Supplement Date Received by Manufacturer05/29/2018
Initial Report FDA Received Date05/10/2018
Supplement Report FDA Received Date06/27/2018
Patient Sequence Number1
Patient Age46 YR
-
-