• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CONMED CORPORATION 19.5 X 86 X 1.27MM (.050 INCH) OSCILLATOR BLADE; BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CONMED CORPORATION 19.5 X 86 X 1.27MM (.050 INCH) OSCILLATOR BLADE; BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL Back to Search Results
Model Number 5071-181
Medical Device Problem Code Detachment Of Device Component (1104)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 04/17/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The reported device is being returned to conmed for evaluation.A supplemental and final report will be filed following the completion of the device evaluation and complaint investigation.
 
Event or Problem Description
The customer reported that the 00507118100, 19.5 x 86 x 1.27 mm (.050 inch) oscillator blade, qty 5, broke during a total knee arthroplasty procedure on (b)(6) 2018.This occurred while the surgeon was cutting the femur with a guide.The breakage was removed from the surgical site causing a slight delay in the procedure.It is reported that there was not any surgical intervention required for this incident and neither the patient nor the user sustained any injury caused by the breakage.This report is being raised on the basis of malfunction with potential for injury upon reoccurrence.
 
Additional Manufacturer Narrative
The alleged broken 00507118100, blade, is not expected to be returned for evaluation and review.This complaint of broken device is unable to be verified and a root cause cannot be determined.The manufacturing documents from the device history record have been reviewed with special attention to the manufacturing and inspection of the product.The product released for distribution were found to have met all specifications prior to shipment.There have been 2 complaints total for this lot number and failure mode within the past two-years.A two-year review of complaint history revealed there has been a total of 2 complaints, regarding 2 devices, for this device family and failure mode.During this same time frame (b)(4) devices have been manufactured and shipped worldwide.(b)(4).Per the instructions for use, the user is advised the following: warnings: do not use the equipment if, upon receipt, the package is opened, damaged, or shows any signs of tampering.Precautions: refer to the appropriate conmed linvatec handpiece instruction manual for device specific warnings.This issue will continue to be monitored through the complaint system to assure patient safety.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
19.5 X 86 X 1.27MM (.050 INCH) OSCILLATOR BLADE
Common Device Name
BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL
Manufacturer (Section D)
CONMED CORPORATION
11311 concept blvd
largo FL 33773
MDR Report Key7502120
Report Number1017294-2018-00060
Device Sequence Number1698495
Product Code GFA
UDI-Device Identifier10845854009086
UDI-Public(01)10845854009086(17)230319(10)916757
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/01/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date03/19/2023
Device Model Number5071-181
Device Catalogue Number00507118100
Device Lot Number916757
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 04/18/2018
Supplement Date Received by Manufacturer05/16/2018
Initial Report FDA Received Date05/10/2018
Supplement Report FDA Received Date06/01/2018
Patient Sequence Number1
-
-