MEDTRONIC PUERTO RICO OPERATIONS CO. VIVA QUAD S; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO Back to Search Results
Model Number DTBB1Q1
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 04/23/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Concomitant product(s): a 4298 lead, implanted: (b)(6) 2018.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the cardiac resynchronization therapy defibrillator (crt-d) system was explanted - 11 days after a routine device replacement procedure - and later replaced due to an infection.No further patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was later reported that after the explant and before the replacement, a temporary pacing lead was used while antibiotic therapy was given.
 
Additional Manufacturer Narrative
Product event summary : the device was returned and analyzed.Returned product analysis was performed and no anomalies were found.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIVA QUAD S
Common Device Name
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
MDR Report Key7502163
Report Number3004209178-2018-10533
Device Sequence Number217349
Product Code NIK
UDI-Device Identifier00643169530201
UDI-Public00643169530201
Combination Product (Y/N)N
Initial Reporter StateOK
Initial Reporter CountryUS
PMA/510(K) Number
P010031
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 06/28/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date10/14/2018
Device Model NumberDTBB1Q1
Device Catalogue NumberDTBB1Q1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/05/2018
Initial Date Received by Manufacturer 04/23/2018
Supplement Date Received by Manufacturer04/30/2018
06/27/2018
Initial Report FDA Received Date05/10/2018
Supplement Report FDA Received Date05/11/2018
06/28/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
6947 LEAD, 5076 LEAD
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age73 YR
-
-