Patient demographics (age at time of event, date of birth, gender, weight) were requested but not provided.No further patient information was provided at the time of this report or made available in response to follow-up communication.No additional adverse consequences have been reported from this event.This device is used for treatment.The device was received and an evaluation was conducted.The complaint was confirmed.Device history record review revealed nothing relevant to this event.The evaluation revealed that the screw's distal tip was broken-off.Complainant's event typically caused by not inserting the implant co-axial to the bone tunnel, prying/leveraging the implant, flexing the joint during insertion or improper bone preparation.The potential causes of this event are being communicated to the event reporter.If additional relevant information is received, a follow-up report will be submitted.
|