• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH GMRS DIST FEM COMP SML L 65MM; PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH GMRS DIST FEM COMP SML L 65MM; PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Catalog Number 64952010
Medical Device Problem Code Device Contamination with Chemical or Other Material (2944)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 04/14/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other events for the reported lot.If additional information is received, it will be provided in a supplemental report upon completion of the investigation.
 
Event or Problem Description
Primary procedure, left femur.It was reported that in opening the implant, a fine white powder was witnessed on the implant.Due to sterility concerns, the implant was not used.Another implant was secured and opened while the cement was setting, with no surgical delay.
 
Additional Manufacturer Narrative
Device evaluated by mfg.An event regarding white powder on implant upon opening the packaging (foreign matter) for a gmrs distal femoral component was reported.The event was not confirmed.Method & results: device evaluation and results: some photographs of the device were provided by the sales rep; there is what appears to be a white powder on the edge of the taper opening on the distal femoral component.The product/pack was returned by the customer.The unit carton containing the distal femoral component (loose in the box), the ifu booklet and the patient labels were returned for evaluation.The returned device appears unremarkable; there is no evidence of the white powder as depicted in the photograph provided.Medical records received and evaluation: not performed as no medical records were provided for medical review.Device history review: indicated all devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the reported lot.Conclusion: some photographs of the device were provided by the sales rep; there is what appears to be a white powder on the edge of the taper opening on the distal femoral component.The product/pack was returned by the customer.The unit carton containing the distal femoral component (loose in the box), the ifu booklet and the patient labels were returned for evaluation.The returned device appears unremarkable; there is no evidence of the white powder as depicted in the photographs provided.Further information such as analysis and identification of the foreign matter, and the return of all packaging components is required to complete the investigation for determining a root cause.No further investigation is required at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
 
Event or Problem Description
Primary procedure, left femur.It was reported that in opening the implant, a fine white powder was witnessed on the implant.Due to sterility concerns, the implant was not used.Another implant was secured and opened while the cement was setting, with no surgical delay.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GMRS DIST FEM COMP SML L 65MM
Common Device Name
PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key7502468
Report Number0002249697-2018-01393
Device Sequence Number19276478
Product Code KRO
UDI-Device Identifier07613327048797
UDI-Public07613327048797
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K023087
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,other
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2022
Device Catalogue Number64952010
Device Lot NumberB9Y9B
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/27/2018
Initial Date Received by Manufacturer 04/14/2018
Supplement Date Received by Manufacturer05/30/2018
Initial Report FDA Received Date05/10/2018
Supplement Report FDA Received Date06/26/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-