• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC SAFE-T-J ROSEN CATHETER EXCHANGE WIRE GUIDE; DQX WIRE, GUIDE, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC SAFE-T-J ROSEN CATHETER EXCHANGE WIRE GUIDE; DQX WIRE, GUIDE, CATHETER Back to Search Results
Model Number G01253
Medical Device Problem Code Device Contamination with Chemical or Other Material (2944)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 04/20/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product code: dqx.(b)(4).This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.
 
Event or Problem Description
It was reported that a hair was visualized inside the package of the unopened safe-t-j rosen catheter exchange wire guide.The device was not used.There was no patient involvement.
 
Additional Manufacturer Narrative
Investigation ¿ evaluation: a review of the complaint history, device history record, documentation, quality control, specifications, and a visual inspection of the returned device were conducted during the investigation.The visual inspection of the returned device confirmed the presence of a black, hair like fiber inside sealed package.The fiber was measured to be 3.9 cm in length.Additionally, a document based investigation evaluation was performed.A review of the device history record showed the lot was released meeting all finished goods release criteria.It should be noted there were no other reported complaints for this lot number.Based on the information provided, the examination of the returned product, and the results of our investigation, the root cause was determined to be related to the device manufacturing process.We will continue to monitor for similar complaints.Per the quality engineering risk assessment, no further action is required.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SAFE-T-J ROSEN CATHETER EXCHANGE WIRE GUIDE
Common Device Name
DQX WIRE, GUIDE, CATHETER
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key7502492
Report Number1820334-2018-01359
Device Sequence Number19146048
Product Code DQX
UDI-Device Identifier00827002012530
UDI-Public(01)00827002012530(17)221213(10)8398537
Combination Product (Y/N)N
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberG01253
Device Catalogue NumberTHSCF-35-260-1.5-ROSEN
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 04/20/2018
Supplement Date Received by Manufacturer07/18/2018
Initial Report FDA Received Date05/10/2018
Supplement Report FDA Received Date07/30/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-