• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC FLEXOR INTRODUCER SHEATH; DYB INTRODUCER, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC FLEXOR INTRODUCER SHEATH; DYB INTRODUCER, CATHETER Back to Search Results
Model Number G07878
Medical Device Problem Code Material Frayed (1262)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 04/23/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(6).(b)(4).Pma/510(k) number: pre-amendment.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.
 
Event or Problem Description
It was reported that prior to a stentgraft placement, the dilator tip of the flexor introducer sheath was found to be frayed.Another device was used instead to complete the procedure.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Additional Manufacturer Narrative
Investigation ¿ evaluation: a review of the complaint history, device history record, drawing, manufacturing instructions, quality control, and a visual inspection and functional evaluation of the returned device were conducted during the investigation.The visual inspection of the returned device confirmed that, although the dilator was determined to have a smooth transition with the wire guide, the distal end was out of round and compressed, with a jagged edge.Additionally, a document based investigation evaluation was performed.There is no evidence to suggest the product was not made to specifications.A review of the device history record showed no nonconforming events which could contribute to this failure mode.It should be noted there were no other reported complaints for this lot number.Based on the information provided, the examination of the returned product, and the results of our investigation, a definitive root cause could not be determined.We will continue to monitor for similar complaints.Per the quality engineering risk assessment, no further action is required.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Additional Manufacturer Narrative
Correction: d1, d4, g5, h4.This mdr is being submitted as having information not previously reported.Additional complaint investigation and record remediation was not performed.Blank fields on this form indicate the information is unknown or unavailable.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXOR INTRODUCER SHEATH
Common Device Name
DYB INTRODUCER, CATHETER
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key7502875
Report Number1820334-2018-01281
Device Sequence Number13866011
Product Code DYB
UDI-Device Identifier00827002078789
UDI-Public(01)00827002078789(17)201127(10)8396651
Combination Product (Y/N)N
PMA/510(K) Number
K142829
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/27/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/27/2020
Device Model NumberG07878
Device Catalogue NumberKCFW-8.0-38-30-RB
Device Lot Number8396651
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 04/24/2018
Supplement Date Received by Manufacturer06/22/2018
03/26/2020
Initial Report FDA Received Date05/10/2018
Supplement Report FDA Received Date07/10/2018
03/27/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-