COOK INC FLEXOR HIGH-FLEX ANSEL GUIDING SHEATH; DYB INTRODUCER, CATHETER Back to Search Results
Model Number G56232
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 04/26/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Pma/510(k) number: pre-amendment.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.
 
Event or Problem Description
It was reported that at the end of a lower extremity angiogram procedure, the check-flo hub broke off the flexor high-flex ansel guiding sheath.The complaint device was introduced through the right common femoral artery in the ipselateral approach and was curved and taped down to the drape.Another manufacturer's 7mm x 250mm stent was used through the flexor high-flex ansel guiding sheath.After removal, another flexor high-flex ansel guiding sheath was used to successfully complete the procedure.According to the initial reporter, a section of the device did not remain in the patients anatomy.No adverse effects were reported as a result of this occurrence.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Additional Manufacturer Narrative
Investigation ¿ evaluation: a review of the dimensional verification, complaint history, device history record, drawing, manufacturing instructions, quality control, and a visual inspection of the returned device were conducted during the investigation.The visual inspection of the returned device confirmed that the sheath tubing was separated from the hub.A 2mm compression was also present in the sheath tubing, beginning at a distance of 8 mm from the proximal end.The inner diameter of the connecting cap was determined to be within specification, while the sheath flare did not pass examination with the flare gauge as it would not pass through either the large lumen or the small lumen.Additionally, a document based investigation evaluation was performed.There is no evidence to suggest the product was not made to specifications.A review of the device history record showed no nonconforming events which could contribute to this failure mode.It should be noted there were no other reported complaints for this lot number.Based on the information provided, the examination of the returned product, and the results of our investigation, a definitive root cause could not be determined.Measures are being conducted to address this failure mode.We will continue to monitor for similar complaints.Per the quality engineering risk assessment, no further action is required.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Additional Manufacturer Narrative
This mdr is being submitted as having information not previously reported.Additional complaint investigation and record remediation was not performed.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXOR HIGH-FLEX ANSEL GUIDING SHEATH
Common Device Name
DYB INTRODUCER, CATHETER
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key7503128
Report Number1820334-2018-01299
Device Sequence Number1652650
Product Code DYB
UDI-Device Identifier00827002562325
UDI-Public(01)00827002562325(17)200329(10)7787165
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K142829
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/29/2020
Device Model NumberG56232
Device Catalogue NumberKCFW-7.0-35-55-RB-HFANL0-HC
Device Lot Number7787165
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 04/26/2018
Supplement Date Received by Manufacturer06/15/2018
03/26/2020
Initial Report FDA Received Date05/10/2018
Supplement Report FDA Received Date06/22/2018
03/30/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
VIBAHN STENT
Patient Age87 YR
-
-