The user facility reported during set up, they noticed a cracked or slice in the purge line that comes off at the top of the fiber bundle.The oxygenator was changed out and the event did not delay beginning or continuing the surgery, the surgery was completed successfully.There was no blood loss due to the event.
|
|
This report is being submitted as follow up no.1 to provide the completed investigation.The actual sample was received by ashitaka for evaluation.Visual inspection revealed that the purge line tube had almost fractured at the joint with the oxygenator purge line port.There were no other anomalies.Magnifying and electron microscopic inspections of the fracture cross- section of the purge line tube on the port side confirmed there was no any anomaly on the surface that could have been a trigger of the generation of the fracture with no embedded foreign substance or entrainment of air bubbles in the material.The surface was in the smooth state on some part and in the rough state on the other part.On the smooth part, some streaks were found to have been generated.This streak pattern implies that the fracture occurred at one point and developed in the direction of the streak pattern.The state of the fracture on the actual sample is experientially known to be consistent with the state when the tube has been damaged as a result of exposition to some shock force under a cold temperature.Review of the temperatures in the u.S.A.From january 2018 when the actual sample was manufactured to (b)(6) 2018 when this complaint occurred found that the lowest temperature had gone down to below zero in some days.With this information being taken into account, the reproductive test was performed with the test conditions being set arbitrarily.A pack of a test product sample which had been cooled down to minus 10 degree celsius beforehand was dropped from 1.5m high.As the result, the tube got fractured at the joint of the port.Based on the state of the fracture surface of the actual sample and the reproductive test result, it is likely the cause of reported event, is that the actual sample was cooled down due to a low temperature during transportation and/or due to the storage environment and that, in this state, it was subjected to some excessive shock force due to being handled inattentively during transportation or storage, resulting in the reported fracture of the tube.The exact cause of the reported event cannot be definitively determined based on the available information.Terumo is continuing to investigate the reported event and the occurrence of this nature, and is instructing the involved carriers and the warehouse staff to handle the product with the best care.
|