Model Number 8637-20
Device Problems Electromagnetic Interference (1194); Volume Accuracy Problem (1675)
Patient Problem Malaise (2359)
Event Date 04/08/2018
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a device manufacturer representative regarding a patient who was receiving morphine (10 mg/ml at 1. 897 mg/day) via an implantable infusion pump for non-malignant pain. It was reported that the patient had not been feeling well since (b)(6) 2018. A motor stall occurred, which was active at the time of the initial report. A motor stall was noted in the attention dialogue box and pump logs. The patient was not hearing audible alarms and when the healthcare provider (hcp) refilled the pump on (b)(6) 2018, there was a notable volume discrepancy. It was unknown if the patient had an mri in (b)(6) 2018. Specifically, the event date was noted to be (b)(6) 2018. They planned to read the pump logs again and find out for certain whether or not the patient had an mri in (b)(6) 2018. It was further reported that it was confirmed that the pump was in a "telemetry state" and a recovery was noted on (b)(6) 2018. The issue was resolved. No further complications were reported.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Additional information was received. It was reported that there was no volume discrepancy noted. No further complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameSYNCHROMED II
Type of DevicePUMP, INFUSION, IMPLANTED, PROGRAMMABLE
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
lisa woodward clark
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263920
MDR Report Key7503223
MDR Text Key107883857
Report Number3004209178-2018-10574
Device Sequence Number1
Product Code LKK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P860004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation
Type of Report Initial,Followup
Report Date 05/21/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received05/10/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date06/14/2018
Device Model Number8637-20
Device Catalogue Number8637-20
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received05/18/2018
Was Device Evaluated by Manufacturer? No Answer Provided
Date Device Manufactured12/15/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-